SafeFAST Top class II A1/A2 Microbiological Safety Cabinets belong to the latest generation of laminar airflow systems manufactured by Faster, in which the choice of materials of construction of the highest quality guarantees conformity to the strictest safety standards.
Faster SafeFAST Top are Microbiological Safety Cabinets – designed and built to performance requirements of the EN-12469:2000 European Standard and NSF/ANSI 49 American Standard, with 70% of the air re-circulated via the main Class H14 HEPA/ULPA filter within the cabinet, whilst the remaining 30% is discharged through an exhaust Class H14 HEPA/ULPA filter.
External structure in epoxy powder coated cold-rolled steel for excellent corrosion resistance to the attack by aggressive common chemicals. Alternatively special models with external structure made in AISI 304L stainless steel for superior cleanability are also available on demand.
Rear wall in stainless steel AISI 304 L, designed to conform to requirements and pass the “cleanability test” according to EN12469:2000.
Work surface in stainless steel AISI 316L consisting of sections (or in one piece upon request) which are easily removable for carrying out routine cleaning and/or autoclaving sterilization procedures; closed or perforated on request.
Work chamber in single piece AISI 304 stainless steel for the best cleanibility.
Front window: Stratified hinged safety glass to give easy access to large items. It is provided with gas springs to keep it open during maintenance or sanitization operations.
Re-circulating and extractor fans: the ‘S’-Series are supplied with single centrifugal fan, whilst models of the ‘D’ Series with double centrifugal fan ( main and exhaust ) to provide complete operator, product and environmental protection. Moreover the ‘D’ models fitted with double motor-fan are designed and are also suitable to discharge the filtered air outside the laboratory through a ducting system if required.
Filtration: H14 HEPA/ULPA filters with an efficiency better than 99,995 % MPPS (EN-1822).
Operation Condition: Air cleanliness in Class ISO 3 as per ISO:EN 14644-1.
The user-friendly practical keyboard and the rear-lit LCD will continuously display all required data keeping the user constantly informed of the cabinet conditions in operation, and in particular:
• display of laminar airflow velocity and frontal air barrier velocity
• display of inside and outside temperature
• display of residual lifetime of HEPA/ULPA filters, UV Lamp and activated carbon filter (if fitted)
• display of total number of hours of operation
• display of saturation level of HEPA/ULPA filters.
Audio-visual alarms provided for:
• out of range or incorrect laminar airflow velocity and frontal air barrier velocity
• front window opened
• clogging of HEPA/ULPA filters
• end of life-cycle of UV lamp and saturation of activated carbon filter (if fitted)
• blockage in the exhaust duct ( if fitted )
• fan-motor malfunction
• power failure
Lighting: fluorescent tubes in built-in housing, placed outside the contaminated area.
Aluminium front closure panel that ensures the sealing of the cabinet for sterilization/fumigation.
D.O.P.-DEHS inlet port for testing the HEPA/ULPA filters.
UV sterilizing lamp (optional) installed on the aluminium front closure panel complete two switch-off countdown timers, one variable on a 0-3 hours scale (1 minute steps), the other set to 3 fixed hours.
Direct Ducting: As standard supplied with 200 mm diameter collar on the top of the cabinet for optional direct ducting to facility exhaust system
SafeFAST Elite Class II B2 Microbiological Safety Cabinets belong to the latest generation of laminar airflow systems manufactured by Faster, in which the choice of materials of construction of the highest quality guarantees conformity to the strictest safety standards.
Faster SafeFAST Elite Class II B2 are Microbiological Safety Cabinets with 100% of the air exhausted as per NSF/ANSI 49; they are designed to provide operator, product and environmental protection and to be vented outside the building without recirculatio
External structure in epoxy powder coated cold-rolled steel for excellent corrosion resistance to the attack by aggressive common chemicals. Alternatively special models with external structure made in AISI 304L stainless steel for superior cleanability are also available on demand.
Rear wall in stainless steel AISI 304 L, designed to conform to requirements and pass the “cleanability test” according to EN12469:2000.
Work surface in stainless steel AISI 316L as standard supplied with spill retaining solid work surface (perforated upon request) which is easily removable for carrying out routine cleaning and/or autoclaving sterilization procedures. Special sloping profile below work surface to ease H14 HEPA/ULPA filter replacement and spilled liquids collection.
Front window: electrically operated vertically sliding safety glass sash window.
System of two motor fans: the units are supplied with double motor-fan in order to discharge the total volume of treated filtered air outside the laboratory through a ducting system.
Filtration: H14 HEPA/ULPA filters with an efficiency better than 99,995 % MPPS (EN-1822).
Prefiltration: Inlet G3 pre-filter efficiency 80≤AM≤90 according to EN 779 and UNI 10339.
Operation Condition: Air cleanliness in Class ISO 3 as per ISO:EN 14644-1.
The user-friendly practical keyboard and the rear-lit LCD will continuously display all required data keeping the user constantly informed of the cabinet conditions in operation, and in particular:
Audio-visual alarms provided for:
Lighting: fluorescent tubes in built-in housing, placed outside the contaminated area
D.O.P.-DEHS inlet port for testing the HEPA/ULPA filters
Exhaust hard duct connection: due to their intrinsic feature to exhaust all air treated, SafeFAST Elite Total Exhaust are supplied with a 300mm diameter collar on top of the unit for direct connection to exhaust system.
Magnetic and removable UV sterilizing lamp (optional) that can be easily placed in each area of the back wall. It is completed with two switch-off countdown timers, one variable on a 0-3 hours scale (1 minute steps), the other set to 3 fixed hours.
SafeFAST Light are the new concept of compact class II A1/A2 Microbiological Safety Cabinet manufactured by Faster, in which the choice of materials of construction of the highest quality guarantees conformity to the strictest safety standards.
Faster SafeFAST Light are Microbiological Safety Cabinets – designed and built to performance requirements of the EN-12469:2000 European Standard and NSF/ANSI 49 American Standard, with 70% of the air re-circulated via the main Class H14 HEPA/ULPA filter within the cabinet, whilst the remaining 30% is discharged through an exhaust Class H14 HEPA/ULPA filter.
External structure in epoxy powder coated cold-rolled steel for excellent corrosion resistance to the attack by aggressive common chemicals. Alternatively special models with external structure made in AISI 304L stainless steel for superior cleanability are also available on demand.
Rear wall in stainless steel AISI 304 L, designed to conform to requirements and pass the “cleanability test”.
Work surface in AISI 304L which is easily removable for carrying out routine cleaning and/or autoclaving sterilization procedures; closed or perforated upon request.
Front window: Stratified hinged safety glass to give easy access to large items. It is provided with gas springs to keep it open during maintenance or sanitization operations.
Re-circulating and extractor fan: SafeFAST Light are supplied with single centrifugal fan.
Servicing side glass with ports for service taps installation.
Filtration: H14 HEPA/ULPA filters with an efficiency better than 99,995 % MPPS (EN-1822).
Operation Condition: Air cleanliness in Class ISO 3 as per ISO:EN 14644-1.
The user-friendly practical keyboard and the rear-lit LCD will continuously display all required data keeping the user constantly informed of the cabinet conditions in operation, and in particular:
Audio-visual alarms provided for:
Lighting: fluorescent tubes in built-in housing, placed outside the contaminated area
D.O.P.-DEHS inlet port for testing the HEPA/ULPA filters
Aluminum front closure panel that ensures the sealing of the cabinet for sterilization/fumigation.
UV sterilizing lamp (optional) installed on the aluminum front closure panel complete two switch-off countdown timers, one variable on a 0-3 hours scale (1 minute steps), the other set to 3 fixed hours.
FlowFAST H are positive pressure ISO Class 3 horizontal laminar flow cabinets which represent far more economical and valid alternatives to clean-rooms, as they are capable of providing uncontaminated, particle-free and sterile conditions for materials in the work-space. They guarantee and maintain excellent product protection factors by offering a positive pressure particle-free working area by virtue of the G3 prefilter with a minimum efficiency of 80%-90% ASHRAE and the H 14 HEPA/ULPA filter with an efficiency better than 99,995 % MPPS (H14 – EN:1822).
External structure in epoxy powder coated cold-rolled steel for excellent corrosion resistance to the attack by aggressive common chemicals. Alternatively special models with external structure made in AISI 304L stainless steel for superior cleanability are also available on demand.
Work surface in stainless steel AISI 304L fixed in one piece.
Filtration:H14 HEPA/ULPA filter with an efficiency better than 99,995 % MPPS (EN-1822).
Prefiltration: Inlet G3 pre-filter efficiency 80≤AM≤90 according to EN 779 and UNI 10339
Operation Condition: Air cleanliness in Class ISO 3 as per ISO:EN 14644-1.
Motor blower: direct coupled motor, electronic speed controlled to maintain a constant laminar air flow of 0.45 m/sec, and compensate for a partially clogged filter up to a maximum plenum pressure of 40 mm of water.
Servicing side glass with ports for service taps installation.
The user-friendly practical keyboard and the rear-lit LCD (optional) will continuously display all required data keeping the user constantly informed of the cabinet conditions in operation, and in particular:
Audio-visual alarms (optional) provided for:
Lighting: fluorescent tubes in built-in housing, placed outside the sterile area.
D.O.P.-DEHS inlet port for testing the HEPA/ULPA filters
UV sterilizing lamp (optional) complete with a flexible front shutter in order to close the working area from dust and for the safety of operators.
ChemFAST Classic are total exhaust chemical fume cupboards working at Constant Air Volume technology (CAV). The inflow velocity can be adjusted from 0,3 to 0,7 m/sec, both during the installation and ordinary maintenance; this feature makes it possible to choose the safest velocities for the removal of contaminants.
ChemFAST Classic complies with the requirements of the EN 14175 part 1, 2, 3 “Machinery Directive 2006/42/ CE” standards designed to guarantee operator and environmental safety with a ventilation outside the building without internal recirculation
Antibacterial coating: Dupont ™ ALESTA® anti-bacterial is able to prevent microbial contamination of surfaces inhibiting in that way the long-term growth of surface contamination.
Compartments under the hood. It is possible to install safety storage cabinets under the hood, since the central space is not taken up by something else, or you can opt for a different solution and have classic or vented cabinets with wheels that are easy to open and close.
Electrical, hydraulic and gas utilities are concentrated on one shoulder of the hood. The electric utilities can be placed inside the hood, as we offer IP66 plug-in sockets, increasing the capacity from 4 sockets to 8 sockets (4 internal and 4 external). Up to 8 controls or alternatively 4 pressure reducers can be installed.
Automatic safety fittings for gas and vacuum and standard electrical sockets in each model and for each size. Internal and external IP 66 protection.
Lighting: the positioning and sizing of the side safety windows provide the highest level of brightness to the work area and create an ideal lighting system.
FILTRATIONS (Optional)
ChemFAST Classic hoods do not have filters as standard, but can be supplied with external filtering groups downstream of the ducted system.
The filters provided can be charcoal or HEPA depending on the type of application that is carried out inside the hood.
Your dedicated partner in providing high-quality laboratory and research equipment. With years of industry experience, we specialize in supplying cutting-edge laboratory instruments, research equipment, and essential lab tools to support scientific discovery, educational advancement, and innovation across various industries.
© 2025 | REIDIUS TECHNOLOGIES | All Rights reserved.